FDA Adverse Event Malfunction Summary report: N

EASYPUMP

MDR report key: 2043201 · Received February 24, 2011

Report

Report Number
2026095-2011-00060
Event Type
Malfunction
Date Received
February 24, 2011
Date of Event
January 6, 2011
Report Date
January 27, 2011
Manufacturer
I-FLOW CORPORATION
Product Code
MEB
PMA / PMN Number
K052117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED FILL VOLUME FOR THIS PUMP WAS 280ML THE DIRECTIONS FOR USE (DFU) (1303046, REV. H), "CAUTIONS: FILLING THE PUMP MORE THAN NOMINAL RESULTS IN SLOWER FLOW RATE." A DEVICE HISTORY RECORD WAS CONDUCTED, AND ALL MANUFACTURING OPERATIONS WERE FOUND TO BE WITHIN SPECIFICATION. CONCLUSION: A FOLLOW-UP REPORT WILL BE FILED WHEN INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

(DRUG/DILUENT: FORTUM 4G) (CATHPLACE: PERIPHERAL CATHETER) FAST FLOW. PUMP INFUSED IN 12 HOURS INSTEAD OF 24 HOURS. (FILL VOLUME: 280ML AND FLOW RATE: 10ML/HR). NO ADVERSE EVENT REPORTED. DATE OF EVENT: (B)(6) 2011. PER DFU: THE NOMINAL FILL VOLUME: 270ML; FLOW RATE: 10ML/HR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYPUMP ELASTOMERIC PUMP MEB I-FLOW CORPORATION LT 270-24 072100

Patients

Seq Age Sex Outcome Treatment
1