FDA Adverse Event Malfunction Summary report: N

TUBE, GASTROINTESTINAL (AND ACCESS)

MDR report key: 2043123 · Received February 24, 2011

Report

Report Number
3006260740-2011-00054
Event Type
Malfunction
Date Received
February 24, 2011
Date of Event
January 31, 2011
Report Date
February 10, 2011
Manufacturer
C. R. BARD, INC. (BASD)
Product Code
KNT
PMA / PMN Number
K915837
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT OF A BREAK IN THE RETENTION DOME IS CONFIRMED, USER RELATED. UPON RECEIPT, A PARTIAL SEMI-CIRCULAR BREAK IN THE RETENTION DOME WAS OBSERVED. A SMALL SECTION OF THE BROKEN DOME IS STILL ATTACHED TO THE FEEDING TUBE AND REVEALS A COMPLETE BOND. THE DULL AND ROUNDED VENEER IDENTIFIED AT THE DOME SITE ARE TRAITS THAT ARE INDICATIVE OF EXCESSIVE FORCE THAT WAS APPLIED DURING THE REMOVAL OF THIS DEVICE. THE LACK OF ANY ASSOCIATED DAMAGE SUGGESTS THE TEAR IN THE DOME WAS CAUSED BY STRETCHING THE DOME MATERIAL BEYOND ITS ELASTIC LIMITS DURING REMOVAL. GROSS VISUAL AND MICROSCOPIC EXAMINATIONS SHOW NO EVIDENCE OF A DEFECT OR DEFORMITY RELATED TO THE MANUFACTURING PROCESS. A CHR IS NOT POSSIBLE, AS NO MANUFACTURING LOT NUMBER INFORMATION HAS BEEN PROVIDED BY THE COMPLAINANT.

Description of Event or Problem · 1

A BREAK IN THE RETENTION DOME DURING TRACTION REMOVAL. THE INDWELL TIME WAS 123 DAYS. THE DOME PORTION WAS REMOVED WITH BOWEL MOVEMENT 3 DAYS LATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TUBE, GASTROINTESTINAL (AND ACCESS) KNT C. R. BARD, INC. (BASD) UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention