FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2043036 · Received March 25, 2011

Report

Report Number
9680959-2011-00875
Event Type
Malfunction
Date Received
March 25, 2011
Date of Event
March 7, 2011
Report Date
March 25, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE REP CONDUCTED AN ONSITE INVESTIGATION. THE POWER SUPPLY FUSE WAS REPLACED. THE SYSTEM WAS TESTED AND IS NOW OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE AMPLIFIER WAS OUT OF SERVICE ON THE 7900 SYSTEM. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1