FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE ARTHRITIS PAD
MDR report key: 2043034
·
Received March 15, 2011
Report
- Report Number
- 1811605-2011-00027
- Event Type
- Malfunction
- Date Received
- March 15, 2011
- Date of Event
- February 1, 2010
- Report Date
- March 2, 2011
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
IT WAS REPORTED AN INTERNAL ELECTRICAL COMPONENT SPARKED AND BURNED THE FABRIC FOUNDATION OF THE UNIT. EXAMINATION OF THE PAD SHOWS ARCING BETWEEN THE LEADS OF ONE OF THE THERMOSTATS THAT RESULTED IN A 1/4" BURN HOLE IN THE COVER OF THE PAD. OTHER EVIDENCE (THE BUNCHED UP CONDITION OF THE PAD AND THE TAPED-DOWN SAFETY SWITCH) SUGGEST THE UNIT WAS USED IMPROPERLY. WE CONCLUDED THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE OF THE PRODUCT ON THE PART OF THE CONSUMER. THIS PARTICULAR STYLE OF THERMOSTAT WAS CHANGED SEVERAL YEARS AGO TO A DESIGN THAT PREVENTS ARCING IN ABUSE CONDITIONS.
Description of Event or Problem · 1
IT WAS REPORTED AN INTERNAL ELECTRICAL COMPONENT SPARKED AND BURNED THE FABRIC FOUNDATION OF THE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE ARTHRITIS PAD | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT COMPANY | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |