FDA Adverse Event Malfunction Summary report: N

THERMOPHORE ARTHRITIS PAD

MDR report key: 2043034 · Received March 15, 2011

Report

Report Number
1811605-2011-00027
Event Type
Malfunction
Date Received
March 15, 2011
Date of Event
February 1, 2010
Report Date
March 2, 2011
Manufacturer
BATTLE CREEK EQUIPMENT COMPANY
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED AN INTERNAL ELECTRICAL COMPONENT SPARKED AND BURNED THE FABRIC FOUNDATION OF THE UNIT. EXAMINATION OF THE PAD SHOWS ARCING BETWEEN THE LEADS OF ONE OF THE THERMOSTATS THAT RESULTED IN A 1/4" BURN HOLE IN THE COVER OF THE PAD. OTHER EVIDENCE (THE BUNCHED UP CONDITION OF THE PAD AND THE TAPED-DOWN SAFETY SWITCH) SUGGEST THE UNIT WAS USED IMPROPERLY. WE CONCLUDED THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE OF THE PRODUCT ON THE PART OF THE CONSUMER. THIS PARTICULAR STYLE OF THERMOSTAT WAS CHANGED SEVERAL YEARS AGO TO A DESIGN THAT PREVENTS ARCING IN ABUSE CONDITIONS.

Description of Event or Problem · 1

IT WAS REPORTED AN INTERNAL ELECTRICAL COMPONENT SPARKED AND BURNED THE FABRIC FOUNDATION OF THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE ARTHRITIS PAD ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT COMPANY 055

Patients

Seq Age Sex Outcome Treatment
1