FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN CARTRIDGE
MDR report key: 2042896
·
Received March 30, 2011
Report
- Report Number
- 2531779-2011-02087
- Event Type
- Malfunction
- Date Received
- March 30, 2011
- Report Date
- February 28, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Removal / Correction Number
- 2531779-02/25/11-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. ALTHOUGH THE CARTRIDGE WAS NOT RETURNED THERE IS NO FURTHER INVESTIGATION NECESSARY WITH RESPECT TO THE LEAK ISSUE. CARTRIDGES WITH LOT# B201581 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.
Description of Event or Problem · 1
IT WAS REPORTED BY THE PATIENT THAT SHE HAS HAD AN ISSUE WITH HER CARTRIDGE LEAKING IMMEDIATELY AFTER FILLING. SHE STATED THAT THE CARTRIDGE WOULD LEAK INSULIN AROUND THE PLUNGER PIECE. THE PATIENT REPORTED SHE DID NOT USE ANY OF THE LEAKING CARTRIDGES, THEREFORE, SHE HAD NO BLOOD GLUCOSE EXCURSIONS.