FDA Adverse Event
Malfunction
Summary report: N
9600
MDR report key: 2042891
·
Received March 25, 2011
Report
- Report Number
- 1720753-2011-02807
- Event Type
- Malfunction
- Date Received
- March 25, 2011
- Date of Event
- March 17, 2011
- Report Date
- March 25, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE X-RAY TUBE, TEMPERATURE SENSOR, AND THE CONNECTORS WERE CLEANED AND RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9600 SYSTEM HAD A TEMPERATURE SENSOR ERROR AND WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.