ANIMAS INSULIN INFUSION PUMP
MDR report key: 2042879
·
Received March 30, 2011
Report
- Report Number
- 2531779-2011-02080
- Event Type
- Malfunction
- Date Received
- March 30, 2011
- Report Date
- March 2, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. DURING INVESTIGATION IT WAS FOUND THAT ALL THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP ARROW KEY CONTACT WAS INVERTED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
PER DISTRIBUTOR, IT WAS REPORTED THAT THE BUTTONS ARE NOT REACTING.