EVERCROSS¿ .035 PTA BALLOON
Report
- Report Number
- 2183870-2011-00059
- Event Type
- Injury
- Date Received
- April 7, 2011
- Date of Event
- March 11, 2011
- Report Date
- March 11, 2011
- Manufacturer
- EV3 INC.
- Product Code
- DQY
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- PHYSICIAN
Narratives
A REVIEW OF THE MANUFACTURING RECORDS FOR THIS DEVICE DID NOT REVEAL ANY DISCREPANCIES RELEVANT TO THIS REPORTED EVENT.
THE PHYSICIAN USED A 6F SHEATH IN THE PATIENT'S ARM AND A .035 WIRE WAS INSERTED THROUGH THE AV FISTULA, CROSSING THE POINT OF STENOSIS THAT NEEDED TO BE ANGIOPLASTIED. THE PHYSICIAN SELECTED A 7X40X80 EVERCROSS PTA BALLOON. THE EVERCROSS BALLOON WAS PREPPED IN THE USUAL MANNER, LOADED OVER THE .035 WIRE, AND CROSSED THROUGH AND POSITIONED WITHIN THE LESION. THE LESION WAS FIBROTIC IN NATURE WITH NO CALCIUM PRESENT. THE PHYSICIAN INFLATED THE BALLOON WITH A 50/50 SOLUTION, HEPARINATED SALINE TO CONTRAST. THE PHYSICIAN THEN INFLATED THE BALLOON TO 10 ATM AND WAS UP AT THE PRESSURE FOR A MAXIMUM OF 10 SECONDS. AFTER DEFLATING THE BALLOON, THE PHYSICIAN TRIED TO RETRACT THE BALLOON OFF THE WIRE, BUT MET WITH RESISTANCE. IN AN EFFORT TO GET THE BALLOON OUT OF THE PATIENT, THE PHYSICIAN HAD TO PUT PRESSURE AT THE ACCESS SITE AND PULL VERY HARD ON THE EVERCROSS BALLOON. EVENTUALLY HE WAS ABLE TO GET THE BALLOON OUT OF THE PATIENT, LOSING WIRE ACCESS. WHEN THE WIRE WAS PULLED OUT OF THE PATIENT, THE BALLOON MATERIAL THAT WAS LEFT ON THE WIRE RIPPED A HOLE THROUGH THE ARTERIAL PATCH AT THE ACCESS POINT. THE PHYSICIAN THEN HAD TO REPAIR THE PATCH WITH ANGIOPLASTY. THE ISSUE WAS RESOLVED AT THIS POINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EVERCROSS¿ .035 PTA BALLOON | CATHETER, PERCUTANEOUS | DQY | EV3 INC. | AB35W07040080 | 9408320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | 6F TERUMO SHEATH| .035 STIFF ANGLED GLIDE WIRE |