FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN CARTRIDGE
MDR report key: 2042848
·
Received March 30, 2011
Report
- Report Number
- 2531779-2011-02075
- Event Type
- Malfunction
- Date Received
- March 30, 2011
- Date of Event
- March 2, 2011
- Report Date
- March 2, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Removal / Correction Number
- 2531779-02/25/11-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. ALTHOUGH THE CARTRIDGE WAS NOT RETURNED, THERE IS NO FURTHER INVESTIGATION NECESSARY WITH RESPECT TO THE LEAK ISSUE. CARTRIDGES WITH LOT# B201581 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT HE HAD A FEW CARTRIDGES THAT LEAKED FROM LOT # B201581. HE STATED THAT HE DID NOT EXPERIENCE ANY BG EXCURSIONS DUE TO THE LEAKING, AND HE DID NOT CALL PREVIOUSLY TO REPORT THE LEAKY CARTRIDGES. THE PATIENT STATED THAT THE PREVIOUSLY USED LEAKING CARTRIDGES ARE NOT AVAILABLE TO BE RETURNED TO ANIMAS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP CARTRIDGE | LZG | ANIMAS CORP. | IR 1200/1250/2020/OTP | B201581 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |