FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2042848 · Received March 30, 2011

Report

Report Number
2531779-2011-02075
Event Type
Malfunction
Date Received
March 30, 2011
Date of Event
March 2, 2011
Report Date
March 2, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K032257
Removal / Correction Number
2531779-02/25/11-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. ALTHOUGH THE CARTRIDGE WAS NOT RETURNED, THERE IS NO FURTHER INVESTIGATION NECESSARY WITH RESPECT TO THE LEAK ISSUE. CARTRIDGES WITH LOT# B201581 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT HE HAD A FEW CARTRIDGES THAT LEAKED FROM LOT # B201581. HE STATED THAT HE DID NOT EXPERIENCE ANY BG EXCURSIONS DUE TO THE LEAKING, AND HE DID NOT CALL PREVIOUSLY TO REPORT THE LEAKY CARTRIDGES. THE PATIENT STATED THAT THE PREVIOUSLY USED LEAKING CARTRIDGES ARE NOT AVAILABLE TO BE RETURNED TO ANIMAS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP CARTRIDGE LZG ANIMAS CORP. IR 1200/1250/2020/OTP B201581

Patients

Seq Age Sex Outcome Treatment
1 50 YR