FDA Adverse Event
Malfunction
Summary report: N
CORE MICRO DRILL
MDR report key: 2042841
·
Received March 30, 2011
Report
- Report Number
- 1811755-2011-01024
- Event Type
- Malfunction
- Date Received
- March 30, 2011
- Date of Event
- May 7, 2009
- Report Date
- March 3, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING QUALITY INVESTIGATION, THE COMPLAINANT'S COMPLAINT WAS CONFIRMED; THE DEVICE OVERHEATED DURING TESTING. FURTHER INVESTIGATION REVEALED CORROSION DAMAGE WITHIN THE INTERNAL PARTS OF THE DEVICE. THE PRIMARY CAUSE OF THE FAILURE WAS ATTRIBUTED TO THE CORRODED MOTOR. THE DAMAGED PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
A STRYKER MICRO DRILL WAS SENT IN FOR REPAIR DUE TO OVERHEATING. THERE WAS NO PT INVOLVEMENT; NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE MICRO DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |