FDA Adverse Event
Malfunction
Summary report: N
NEPTUNE ROVER WITHOUT SMOKE EVAC AND POWER POLE
MDR report key: 2042840
·
Received March 30, 2011
Report
- Report Number
- 1811755-2011-01013
- Event Type
- Malfunction
- Date Received
- March 30, 2011
- Date of Event
- March 1, 2011
- Report Date
- March 1, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- FYD
- PMA / PMN Number
- K012991
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE POWER CORD WAS REPLACED ON THE DEVICE. THE DEVICE THEN MET ALL SPECIFICATION AND WAS PUT BACK INTO SERVICE. OVER EXTENDED YEARS OF USE THE POWER PLUG WIRES CAN DETACH OR BECOME LOOSENED FROM THE POWER PLUG CAUSING LOSS OF SYSTEM FUNCTION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PRONGS ON THE POWER PLUG WERE LOOSE, CAUSING ARCING AND PRODUCING HEAT. THIS WAS DISCOVERED DURING A SERVICE REPAIR. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES.