FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITHOUT SMOKE EVAC AND POWER POLE

MDR report key: 2042840 · Received March 30, 2011

Report

Report Number
1811755-2011-01013
Event Type
Malfunction
Date Received
March 30, 2011
Date of Event
March 1, 2011
Report Date
March 1, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE POWER CORD WAS REPLACED ON THE DEVICE. THE DEVICE THEN MET ALL SPECIFICATION AND WAS PUT BACK INTO SERVICE. OVER EXTENDED YEARS OF USE THE POWER PLUG WIRES CAN DETACH OR BECOME LOOSENED FROM THE POWER PLUG CAUSING LOSS OF SYSTEM FUNCTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PRONGS ON THE POWER PLUG WERE LOOSE, CAUSING ARCING AND PRODUCING HEAT. THIS WAS DISCOVERED DURING A SERVICE REPAIR. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITHOUT SMOKE EVAC AND POWER POLE AIR-HANDLING APPARATUS FOR A SURGICAL OPERATING ROOM FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK