ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-02056
- Event Type
- Malfunction
- Date Received
- March 30, 2011
- Report Date
- March 2, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVAL IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 (B)(4) 2011-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR, DISLODGED DISPLAY SCREEN, AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH WAS DUPLICATED DURING TESTING; DURING TESTING THE LOAD CARTRIDGE STEP DID NOT RECOGNIZE THE CARTRIDGE AND PUSHED FLUID OUT EMITTING A "NO CARTRIDGE DETECTED" WARNING. UNRELATED TO THE COMPLAINT, THE BOLUS BUTTON COVER WAS FOUND TO BE PUNCTURED.
PT REPORTS PUMP DISPENSED INSULIN DURING LOAD CARTRIDGE PHASE.