FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2042828 · Received March 30, 2011

Report

Report Number
2531779-2011-02056
Event Type
Malfunction
Date Received
March 30, 2011
Report Date
March 2, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVAL IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(4) 2011-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR, DISLODGED DISPLAY SCREEN, AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH WAS DUPLICATED DURING TESTING; DURING TESTING THE LOAD CARTRIDGE STEP DID NOT RECOGNIZE THE CARTRIDGE AND PUSHED FLUID OUT EMITTING A "NO CARTRIDGE DETECTED" WARNING. UNRELATED TO THE COMPLAINT, THE BOLUS BUTTON COVER WAS FOUND TO BE PUNCTURED.

Description of Event or Problem · 1

PT REPORTS PUMP DISPENSED INSULIN DURING LOAD CARTRIDGE PHASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1