FDA Adverse Event
Malfunction
Summary report: N
MAESTRO MEDIUM FIXED DURAGUARD
MDR report key: 2042802
·
Received March 30, 2011
Report
- Report Number
- 1811755-2011-00977
- Event Type
- Malfunction
- Date Received
- March 30, 2011
- Date of Event
- February 25, 2011
- Report Date
- March 1, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DRILL ATTACHMENT WAS RETURNED TO THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. THE DEVICE EVALUATION CONFIRMED THAT THE ARM OF THE DURAGUARD ATTACHMENT HAD BROKEN OFF AT THE BASE. THE CAUSE OF THE BREAKAGE IS UNKNOWN, BUT IS LIKELY DUE TO EXCESSIVE SIDE FORCE BEING APPLIED TO THE DEVICE DURING USE. THE DRILL ATTACHMENT COULD NOT BE REPAIRED AND THEREFORE, WAS NOT RETURNED TO THE USER FACILITY.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A CRANIOTOMY, THE LEG PORTION OF THE DURAGUARD ATTACHMENT BROKE OFF. NO DEVICE FRAGMENTS ENTERED THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITHOUT A DELAY. NO PATIENT OR USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAESTRO MEDIUM FIXED DURAGUARD | MOTOR, DRILL, PNEUMATIC | HBB | STRYKER INSTRUMENTS KALAMAZOO | 09229 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |