FDA Adverse Event Malfunction Summary report: N

MAESTRO MEDIUM FIXED DURAGUARD

MDR report key: 2042802 · Received March 30, 2011

Report

Report Number
1811755-2011-00977
Event Type
Malfunction
Date Received
March 30, 2011
Date of Event
February 25, 2011
Report Date
March 1, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DRILL ATTACHMENT WAS RETURNED TO THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. THE DEVICE EVALUATION CONFIRMED THAT THE ARM OF THE DURAGUARD ATTACHMENT HAD BROKEN OFF AT THE BASE. THE CAUSE OF THE BREAKAGE IS UNKNOWN, BUT IS LIKELY DUE TO EXCESSIVE SIDE FORCE BEING APPLIED TO THE DEVICE DURING USE. THE DRILL ATTACHMENT COULD NOT BE REPAIRED AND THEREFORE, WAS NOT RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CRANIOTOMY, THE LEG PORTION OF THE DURAGUARD ATTACHMENT BROKE OFF. NO DEVICE FRAGMENTS ENTERED THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITHOUT A DELAY. NO PATIENT OR USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAESTRO MEDIUM FIXED DURAGUARD MOTOR, DRILL, PNEUMATIC HBB STRYKER INSTRUMENTS KALAMAZOO 09229

Patients

Seq Age Sex Outcome Treatment
1 UNK