ACTIVE® INHIBIN A ELISA, 4X96 WELLS
Report
- Report Number
- 2122870-2011-00911
- Event Type
- Malfunction
- Date Received
- April 7, 2011
- Date of Event
- January 24, 2011
- Report Date
- February 18, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- NDR
- PMA / PMN Number
- K002128
- Removal / Correction Number
- 2050012-03/31/2011-014C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
SAMPLE COLLECTION AND CENTRIFUGATION DATA WAS NOT SUPPLIED. SAMPLE STORAGE INCLUDING CONDITIONS FOR TIME FRAME BETWEEN LOT CORRELATION EVALUATIONS WAS NOT SUPPLIED. QC DATA POINTS IDENTIFIED THAT THE QC RECOVERY WAS WITHIN THE ESTABLISHED RANGES. HOWEVER, QC WAS OUT OF RANGE AT THE END OF THE PLATE ASSAY RUN PERFORMED ON (B)(6) 2011. THE CUSTOMER INDICATED THAT THE UNIT ASSOCIATED WITH THIS EVENT WAS PERFORMING WITHIN THE LABORATORY'S SPECIFICATIONS. SYMPTOM REPORTED IS RELATIVE TO CUSTOMER'S EXPECTATIONS OF % DIFFERENCE BETWEEN LOTS WHICH WAS NOT RELATED. INTERNAL INVESTIGATION IS PENDING TO DATE, AND NO CLEAR ROOT CAUSE HAS BEEN DETERMINED THUS FAR.
A CUSTOMER CONTACTED BECKMAN COULTER INC (BCI) STATING THAT THE DATA GENERATED DURING LOT TO LOT CORRELATIONS FOR THE ACTIVE INHIBIN A ELISA METHODOLOGY DID NOT MEET THE LABORATORY'S EXPECTATION CRITERIA OF LESS THAN 10% DIFFERENCE FOR THE SAMPLE COMPARISON. THE CUSTOMER INDICATED THAT THE LOT IS EXHIBITING A POSITIVE BIAS WHEN COMPARED TO OLDER LOT SAMPLES RECOVERIES. THE CUSTOMER ALSO INDICATED THAT THE COMPARISON ASSAY ANALYSIS WAS PERFORMED ON MULTIPLE DAYS AND NOT A TRUE SIDE BY SIDE ANALYSIS. THE ORIGINAL ANALYSIS WAS PERFORMED ON (B)(6) 2011 WITH THE OLD LOT NUMBER, FOLLOWED BY THE COMPARATIVE ANALYSIS ON (B)(6) 2011 RESPECTIVELY WITH THE NEW LOT NUMBER. CORRELATION ANALYSIS DATA REPORT PROVIDED, ILLUSTRATED FIFTY FIVE (55) COMPARISON DATA POINTS ANALYZED ON (B)(6) 2011, DEMONSTRATING A 15% POSITIVE BIAS BETWEEN THE OLD AND THE NEW LOT. ADDITIONAL CORRELATION ANALYSIS DATA REPORT PROVIDED, ILLUSTRATED FIFTY FIVE (55) COMPARISON DATA POINTS ANALYZED ON (B)(6) 2011, DEMONSTRATING A 12% POSITIVE BIAS BETWEEN THE OLD AND THE NEW LOT. THE STUDIES DID NOT INCLUDE ANY PATIENT SAMPLES THAT WERE USED FOR REPORTING. THE CUSTOMER DID NOT REPORT AFFECT TO THE PATIENT OR USER ATTRIBUTED OR CONNECTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACTIVE® INHIBIN A ELISA, 4X96 WELLS | INHIBIN A ENZYME-LINKED IMMUNOSORBENT (ELISA) | NDR | BECKMAN COULTER INC. | NA | 091174 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |