FDA Adverse Event Malfunction Summary report: N

ACTIVE® INHIBIN A ELISA, 4X96 WELLS

MDR report key: 2042765 · Received April 7, 2011

Report

Report Number
2122870-2011-00911
Event Type
Malfunction
Date Received
April 7, 2011
Date of Event
January 24, 2011
Report Date
February 18, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
NDR
PMA / PMN Number
K002128
Removal / Correction Number
2050012-03/31/2011-014C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SAMPLE COLLECTION AND CENTRIFUGATION DATA WAS NOT SUPPLIED. SAMPLE STORAGE INCLUDING CONDITIONS FOR TIME FRAME BETWEEN LOT CORRELATION EVALUATIONS WAS NOT SUPPLIED. QC DATA POINTS IDENTIFIED THAT THE QC RECOVERY WAS WITHIN THE ESTABLISHED RANGES. HOWEVER, QC WAS OUT OF RANGE AT THE END OF THE PLATE ASSAY RUN PERFORMED ON (B)(6) 2011. THE CUSTOMER INDICATED THAT THE UNIT ASSOCIATED WITH THIS EVENT WAS PERFORMING WITHIN THE LABORATORY'S SPECIFICATIONS. SYMPTOM REPORTED IS RELATIVE TO CUSTOMER'S EXPECTATIONS OF % DIFFERENCE BETWEEN LOTS WHICH WAS NOT RELATED. INTERNAL INVESTIGATION IS PENDING TO DATE, AND NO CLEAR ROOT CAUSE HAS BEEN DETERMINED THUS FAR.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC (BCI) STATING THAT THE DATA GENERATED DURING LOT TO LOT CORRELATIONS FOR THE ACTIVE INHIBIN A ELISA METHODOLOGY DID NOT MEET THE LABORATORY'S EXPECTATION CRITERIA OF LESS THAN 10% DIFFERENCE FOR THE SAMPLE COMPARISON. THE CUSTOMER INDICATED THAT THE LOT IS EXHIBITING A POSITIVE BIAS WHEN COMPARED TO OLDER LOT SAMPLES RECOVERIES. THE CUSTOMER ALSO INDICATED THAT THE COMPARISON ASSAY ANALYSIS WAS PERFORMED ON MULTIPLE DAYS AND NOT A TRUE SIDE BY SIDE ANALYSIS. THE ORIGINAL ANALYSIS WAS PERFORMED ON (B)(6) 2011 WITH THE OLD LOT NUMBER, FOLLOWED BY THE COMPARATIVE ANALYSIS ON (B)(6) 2011 RESPECTIVELY WITH THE NEW LOT NUMBER. CORRELATION ANALYSIS DATA REPORT PROVIDED, ILLUSTRATED FIFTY FIVE (55) COMPARISON DATA POINTS ANALYZED ON (B)(6) 2011, DEMONSTRATING A 15% POSITIVE BIAS BETWEEN THE OLD AND THE NEW LOT. ADDITIONAL CORRELATION ANALYSIS DATA REPORT PROVIDED, ILLUSTRATED FIFTY FIVE (55) COMPARISON DATA POINTS ANALYZED ON (B)(6) 2011, DEMONSTRATING A 12% POSITIVE BIAS BETWEEN THE OLD AND THE NEW LOT. THE STUDIES DID NOT INCLUDE ANY PATIENT SAMPLES THAT WERE USED FOR REPORTING. THE CUSTOMER DID NOT REPORT AFFECT TO THE PATIENT OR USER ATTRIBUTED OR CONNECTED TO THIS EVENT.