FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED HUMIDIFICATION CHAMBER

MDR report key: 2042762 · Received April 6, 2011

Report

Report Number
9611451-2011-00219
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
February 22, 2011
Report Date
March 9, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V HUMIDIFICATION CHAMBERS HAVE NOT BEEN RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR INSPECTION. OUR ANALYSIS IS BASED ON THE INFORMATION PROVIDED BY THE HOSPITAL. RESULTS: THE NURSE FROM THE HOSPITAL REPORTED THAT THE LEAK OBSERVED WAS DUE TO THE "HAIRLINE CRACK" IN THE WATER FEEDSET TUBE NEXT TO THE WATERBAG SPIKE. IT WAS FURTHER REPORTED THAT THE WATER FEEDSET TUBE "WAS UNDER TENSION BECAUSE NO FEED SET EXTENSION WAS USED AND THIS RESULTED IN THE CRACK." A LOT CHECK WAS NOT PERFORMED AS NO LOT INFORMATION HAD BEEN PROVIDED. CONCLUSION: WITHOUT THE RETURN OF THE COMPLAINT DEVICE, WE ARE UNABLE TO DETERMINE THE POSSIBLE ROOT CAUSE OF THE REPORTED FAULT. ALL CHAMBERS ARE PRESSURE TESTED BEFORE LEAVING THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. OUR USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THE CONNECTION PORTS OF TWO MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE FOUND CRACKED. THIS WAS OBSERVED BEFORE USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VENTED AUTOFEED HUMIDIFICATION CHAMBER BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED MR290V NOT PROVIDED

Patients

Seq Age Sex Outcome Treatment
1