FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 2042742
·
Received April 6, 2011
Report
- Report Number
- 2050012-2011-00970
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- March 4, 2011
- Report Date
- March 4, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PER CUSTOMER, THE LEAK WAS OBSERVED AT THE GLUM CUP AFTER PERFORMING CUP MAINTENANCE. A BCI FIELD SERVICE ENGINEER (FSE) CLEANED RETAINER AND PORT WITH DI WATER. PUT EXISTING GLUCOSE SENSOR BACK ON ANALYZER. NO LEAKING OBSERVED. FSE PRIMED GLUM AND OBSERVED PROPER FILLING AND STIRRING. CALIBRATED GLUCOSE SENSOR. FSE STATED THAT THE CUSTOMER HAD SMALL AMOUNT OF REAGENT DRIP FROM AROUND RETAINER. FSE CALIBRATED THE GLUCOSE SENSOR AND VERIFIED INSTRUMENT PERFORMANCE PER ESTABLISHED PROCEDURE.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO GLUCOSE CUP LEAK ON THE UNICEL DXC 600I SYNCHRON CHEMISTRY ANALYZER. NO INJURY OR EXPOSURE TO BIOHAZARDOUS MATERIALS WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600I | N/A |