FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2042731 · Received April 6, 2011

Report

Report Number
2050012-2011-00972
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 4, 2011
Report Date
March 4, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PER CUSTOMER, THE ISSUE STARTED ON (B)(6) 2011 AND UPON NOTICING MULTIPLE CALIBRATION FAILURES ON SOME TDM TESTS AND NOTICING LIQUID ON THE BLACK TRAY OF THE REAGENT CAROUSEL. CUSTOMER DRIED UP THE LIQUID ON THE REAGENT CAROUSEL TRAY AND LOADED NEW CHEMISTRY CARTRIDGES (CC); CALIBRATED AND RAN QC. QC READ IN RANGE FOR ALL CHEMISTRIES ON THE CC SIDE. A BCI FIELD SERVICE ENGINEER PERFORMED THE FOLLOWING: REPLACED REAGENT SYRINGE T-VALVE, A/B VALVE, AND BOTH WASH COLLAR VACUUM VALVES. REPLACED VACUUM PUMP DUE TO INTERMITTENT LOW VACUUM ERRORS. PRIMED X 20. THE CUSTOMER RAN FULL QC PANEL, & RESUMED OPERATION WITHOUT ERRORS OR LEAKS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO RECURRING REAGENT CAROUSEL TRAY LEAK ON THE UNICEL DXC 600I SYNCHRON CHEMISTRY ANALYZER. NO INJURY OR EXPOSURE TO BIOHAZARDOUS MATERIALS WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600I N/A

Patients

Seq Age Sex Outcome Treatment
1