FDA Adverse Event Injury Summary report: N

CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS

MDR report key: 2042720 · Received April 6, 2011

Report

Report Number
2015691-2011-15156
Event Type
Injury
Date Received
April 6, 2011
Date of Event
December 8, 2010
Report Date
March 7, 2011
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
PMA / PMN Number
P860057/S001
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE NOT RETURNED. ADDITIONAL MANUFACTURER NARRATIVE: UNFORTUNATELY, THE EDWARDS DEVICE WAS NOT RETURNED FOR ANALYSIS. THE DEVICE HISTORY RECORD (DHR) REVIEW IS CURRENTLY IN PROCESS. CALCIFICATION IS A WELL RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION.

Additional Manufacturer Narrative · 1

ADDITIONAL MANUFACTURER NARRATIVE: THE DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED AND THERE WAS NO NONCONFORMANCE FOUND RELATED TO THIS EVENT.

Description of Event or Problem · 1

EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE. IN THIS CASE, IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AFTER AN IMPLANT DURATION OF APPROXIMATELY 8 1/2 YEARS. THROUGH FOLLOW-UP WITH THE HEALTH-CARE PROVIDER, IT WAS LEARNED THAT THE DEVICE WAS EXPLANTED DUE TO AORTIC STENOSIS. ACCORDING TO THE OPERATIVE REPORT, TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) CONFIRMED SEVERE DEGENERATIVE PROSTHETIC AORTIC VALVE STENOSIS WITH MODERATE AORTIC REGURGITATION SECONDARY TO LEAFLET CONTRACTION AND FIXATION. THERE WAS NO EVIDENCE OF ENDOCARDITIS. UPON REMOVAL, IT WAS NOTED THAT THE VALVE LEAFLETS WERE CALCIFIED. THE VALVE WAS EXCISED AND THE ANNULUS WAS AGGRESSIVELY DEBRIDED OF RESIDUAL PLEDGETS. ANOTHER EDWARDS BIOPROSTHETIC VALVE WAS IMPLANTED. POST-OPERATIVE TEE DEMONSTRATED NORMAL BIOPROSTHETIC LEAFLET MOBILITY, NO EVIDENCE OF PERIVALVULAR LEAK, AND BASELINE VENTRICULAR FUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES 2800 2E0850

Patients

Seq Age Sex Outcome Treatment
1 82 YR Required Intervention