FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM

MDR report key: 2042643 · Received April 6, 2011

Report

Report Number
2134265-2011-01086
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 16, 2011
Report Date
March 16, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE FOUND THE DEVICE WAS RETURNED TO THE COMPLAINT INVESTIGATION SITE WITH PROXIMAL STENT DAMAGE. STRUT ROWS AT THE PROXIMAL END OF THE STENT WERE RAISED AND MISALIGNED. THIS TYPE OF DAMAGE IS CONSISTENT WITH THE STENT MEETING RESISTANCE UPON ADVANCEMENT AND/OR WITHDRAWAL. THE TIP AND BALLOON SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. MANDREL RESISTANCE WAS ENCOUNTERED DUE TO THE PRESENCE OF SOLIDIFIED BLOOD WITHIN THE ENTIRE LENGTH OF THE INFLATION LUMEN, THEREFORE INDICATING THE DEVICE HAD BEEN USED IN VIVO. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4)

Additional Manufacturer Narrative · 1

(B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE STENT DAMAGE OCCURRED. AFTER PREDILATION OF THE RIGHT INTERMEDIUS ARTERY, A 2.25X12MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LESION. IT WAS NOTED THAT THE STENT BECAME DAMAGED WHILE TRYING TO CROSS THE LESION AND THE DEVICE WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE WITH NO PATIENT COMPLICATIONS REPORTED. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE STENT DAMAGE OCCURRED. AFTER PREDILATION OF THE RIGHT INTERMEDIUS ARTERY, A 2.25X12MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LESION. IT WAS NOTED THAT THE STENT BECAME DAMAGED WHILE TRYING TO CROSS THE LESION AND THE DEVICE WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE WITH NO PATIENT COMPLICATIONS REPORTED. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE.