HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-04163
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- March 1, 2011
- Report Date
- March 15, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- NURSE
Narratives
(B)(4). EVALUATION SUMMARY: THIS REPORT FOR A LEAK WAS CONFIRMED IN THE LAB. A VISUAL INSPECTION WAS PERFORMED AND A CRACK WAS NOTED IN THE CORNER END OF THE CASSETTE. THE CASSETTE WAS LOADED ONTO THE HOMECHOICE MACHINE AND A FUNCTIONAL INSPECTION WAS PERFORMED. A BATCH REVIEW WAS PERFORMED AND NO EXCEPTIONS OR NONCONFORMANCE HAVE BEEN FOUND. A ROOT CAUSE WAS NOT IDENTIFIED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER; HOWEVER THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED. A 510K NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510K NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS THE SAME OR SIMILAR TO THE PRODUCT DISTRIBUTED WITHIN THE U.S.
ON (B)(6) 2011, A NURSE CONTACTED BAXTER (B)(4) AND REPORTED THAT WHEN THE PATIENT HAD FILLED THE CASSETTE (FILLED THE TUBES) AND WAS GOING TO START THERAPY, THE TREATMENT SOLUTION HAD LEAKED OUT FROM THE DOOR OF THE HOMECHOICE MACHINE WHERE THE CASSETTE WAS PLACED. THE OCCURRENCE DATE IS UNKNOWN. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE | S10A13108 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HOMECHOICE |