FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 2042608 · Received April 6, 2011

Report

Report Number
3005075853-2011-01389
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 12, 2011
Report Date
March 14, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. NO OPENING ISSUES WERE NOTED DURING EVALUATION; THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE VISUAL INDICATOR WAS OVER-TRAVELED. WHEN THIS OCCURS, THE INDICATOR WHEEL MAY CONTINUE TO ROTATE AFTER INDICATION OF A COMPLETELY FIRED DEVICE. PLEASE NOTE THE OVER-TRAVEL OF THE INDICATOR WHEEL IS NOT RELATED TO THE REPORTED EVENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE THE DEVICE LOCKED ON THE CYSTIC DUCT. THEY REMOVED THE DEVICE BY PULLING ON THE HANDLES APART AND SPRINGING THE JAWS LOOSE. THERE WAS NO DAMAGE TO THE TISSUE REPORTED. THEY COMPLETED THE PROCEDURE WITH A NEW LIKE DEVICE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA H4354K

Patients

Seq Age Sex Outcome Treatment
1