VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM

MDR report key: 2042607 · Received April 6, 2011

Report

Report Number
1319681-2011-00076
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 8, 2011
Report Date
April 6, 2011
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
MMI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION DETERMINED THAT A NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS TROP I ES RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE PROCESSED ON THE VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM. AN OCD FIELD ENGINEER PERFORMED SERVICE ACTIVITIES INCLUDING CLEANING OF THE INCUBATOR. FOLLOWING THESE ACTIONS, ACCEPTABLE VITROS TROP I ES PERFORMANCE WAS OBSERVED. IN ADDITION, THE INVESTIGATION DETERMINED THAT THE SAMPLES IN QUESTION WERE NOT PROCESSED IN ACCORDANCE WITH THE TUBE MANUFACTURER'S RECOMMENDATIONS. IT IS LIKELY THAT CELLULAR DEBRIS, DUE TO POOR SAMPLE PREPARATION, WAS PRESENT IN THE AFFECTED SAMPLES, ALTHOUGH THIS COULD NOT BE CONFIRMED. THE ROOT CAUSE OF THE EVENT IS UNKNOWN. THE POSSIBILITY THAT POOR SAMPLE PROCESSING OR AN ANALYZER RELATED EVENT HAD CONTRIBUTED TO THE RESULTS COULD NOT BE RULED OUT.

Description of Event or Problem · 1

THE CUSTOMER OBTAINED A NON REPRODUCIBLE, HIGHER THAN EXPECTED VITROS TROP I ES RESULT FROM ONE PATIENT SAMPLE WHEN PROCESSED ON THE VITROS ECI Q IMMUNODIAGNOSTIC SYSTEM. BIASED RESULTS OF THE DIRECTION AND MAGNITUDE OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. CONFLICTING INFORMATION WAS PROVIDED IN REGARDS TO WHETHER THE RESULT WAS NOT REPORTED OUT OF THE LABORATORY OR NOT. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITROS ECIQ IMMUNODIAGNOSTIC SYSTEM IMMUNODIAGNOSTIC ANALYZER MMI ORTHO-CLINICAL DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1