STARCLOSE SE VASCULAR CLOSURE SYSTEM
Report
- Report Number
- 2024168-2011-02413
- Event Type
- Injury
- Date Received
- April 6, 2011
- Date of Event
- March 14, 2011
- Report Date
- March 16, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P050007
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
(B)(4). EVALUATION SUMMARY: EVALUATION OF THE RETURNED DEVICE FOUND THAT THE MOVEMENT OF AN INTERNAL COMPONENT (GARAGE BLOCK) WAS RESTRICTED WHICH RESULTED IN THE PROXIMAL DISPLACEMENT OF THE COMPONENTS TUBE AND EXPOSURE OF THE LOADED CLIP. THIS PREVENTED FURTHER DEPLOYMENT OF THE DEVICE AND THE CLIP BY LIMITING THE MOVEMENT OF THE CLIP PUSHER BLOCK/TUBE ASSEMBLY. THE PUSHER BLOCK DID NOT CONNECT WITH A COMPONENT AT THE DISTAL END OF THE RELEASE ROD TO COLLAPSE THE LOCATOR WINGS AND DISENGAGE THE PLUNGER DUE TO THE RESTRICTED MOVEMENT OF THE GARAGE BLOCK. THE CAUSE FOR THE RESTRICTED MOVEMENT OF THE GARAGE BLOCK AND SUBSEQUENT INABILITY TO DEPLOY THE CLIP IS DUE TO RESISTANCE ENCOUNTERED WHILE ADVANCING THE THUMB ADVANCER. RESISTANCE ENCOUNTERED DURING THUMB ADVANCER DEPLOYMENT IS CONSISTENT WITH RADIAL FORCE CONSTRICTION ON THE SHEATH. FACTORS THAT MAY CONTRIBUTE TO RADIAL FORCE CONSTRICTION MAY INCLUDE A TIGHT TISSUE TRACT OR ANATOMICAL CONDITIONS. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT. A QUERY OF THE COMPLAINT DATABASE FOR THIS LOT REVEALED NO SIMILAR INCIDENTS WITH REPORTED PREMATURELY DEPLOYED CLIP. A QUERY OF THE COMPLAINT DATABASE FOR THIS LOT BASED ON ACTUAL INVESTIGATION FINDINGS REVEALED NO SIMILAR INCIDENTS THAT EXHIBITED THE PUSHER-TO-GARAGE BLOCK DISPLACEMENT THAT SUBSEQUENTLY RESULTED IN A FAILURE TO FIRE THE CLIP.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE ATTEMPTED ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, THE CLIP PREMATURELY DEPLOYED IN THE TISSUE TRACK. THE CLIP WAS REMOVED AND HEMOSTASIS WAS ACHIEVED USING MANUAL COMPRESSION. THERE WAS NO ADVERSE PATIENT SEQUELA. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STARCLOSE SE VASCULAR CLOSURE SYSTEM | IMPLANTABLE CLIP | MGB | AV-TEMECULA-CT | 010126H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Required Intervention |