FDA Adverse Event Injury Summary report: N

20" CENTURY SCIENTIFIC VAC

MDR report key: 2042526 · Received April 6, 2011

Report

Report Number
3005899764-2011-00009
Event Type
Injury
Date Received
April 6, 2011
Date of Event
March 10, 2011
Report Date
April 6, 2011
Manufacturer
STERIS MEXICO, S. DE R.L. DE C.V.
Product Code
FLE
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE STERIS SERVICE TECHNICIAN BELIEVED HE HAD SUFFICIENTLY DRAINED ALL RESIDUAL STEAM FROM THE GENERATOR AND PROCEEDED TO SERVICE THE UNIT. IN THE PROCESS OF REMOVING THE HEATING ELEMENT, RESIDUAL STEAM ESCAPED THE UNIT AND IMPACTED THE TECHNICIAN WHO WAS POSITIONED AT THE OPENING OF THE CHAMBER. ON (B)(6), 2011 STERIS CORPORATE HEALTH AND SAFETY ISSUED A NOTIFICATION TO FIELD MANAGEMENT TO REMIND ALL TECHNICIANS OF THE PROPER INSTRUCTIONS FOR PERFORMING PREVENTATIVE MAINTENANCE ON STEAM STERILIZER GENERATORS AND TO ALWAYS FOLLOW ESTABLISHED PROCEDURES WHEN PERFORMING WORK.

Description of Event or Problem · 1

STERIS CONFIRMED THE TECHNICIAN HAS RETURNED TO NORMAL WORK DUTIES.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT A STERIS SERVICE TECHNICIAN WAS PERFORMING PREVENTATIVE MAINTENANCE ON THE HEATING ELEMENT OF THE STERILIZER'S STEAM GENERATOR WHEN HOT STEAM CAME OUT OF THE GENERATOR'S CHAMBER AND ONTO HIS HANDS, ARMS AND LEG. HE IS BEING TREATED FOR BURNS TO HIS HANDS, ARMS AND LEG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 20" CENTURY SCIENTIFIC VAC STERILIZER, STEAM FLE STERIS MEXICO, S. DE R.L. DE C.V.

Patients

Seq Age Sex Outcome Treatment
1 Other