STEAM GENERATOR

MDR report key: 2042524 · Received April 6, 2011

Report

Report Number
3005899764-2011-00010
Event Type
Injury
Date Received
April 6, 2011
Date of Event
March 22, 2011
Report Date
April 6, 2011
Manufacturer
STERIS MEXICO, S. DE R.L. DE C.V.
Product Code
FLE
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN INSPECTED THE GENERATOR AND FOUND THE UNIT WAS OPERATING PROPERLY. THE CAUSE OF THIS EVENT IS OPERATOR ERROR AS THE USER FACILITY REPORTED THE OPERATOR TURNED THE POWER TO THE GENERATOR ON, LEFT THE ROOM AND RETURNED LATER TO OPEN THE DRAIN VALVES TO DRAIN THE UNIT. THE PROPER PROCEDURE IS TO OPEN THE VALVES AND DRAIN THE UNIT WHILE THE POWER IS OFF AND THE UNIT IS COOL. THE EQUIPMENT MAINTENANCE MANUAL STATES CAUTION: "WAIT UNTIL BOILER PRESSURE GAUGE INDICATES "0" PRESSURE AND HAS COOLED TO 140 F BEFORE PERFORMING FLUSH PROCEDURE. POSITION CONTROL POWER SWITCH OFF, OPEN THE BOILER DRAIN VALVE." THE STERIS SERVICE TECHNICIAN PERFORMED AN IN-SERVICE ON (B)(6) 2011 ON THE CORRECT METHOD TO PERFORM THE DRAIN AND FLUSH PROCEDURE ON THE GENERATOR.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT AN OPERATOR OPENED THE VALVES TO PERFORM THE DRAIN AND FLUSH PROCEDURE FOR A GENERATOR WHEN STEAM AND HOT WATER ESCAPED FROM THE UNIT, CAUSING A BURN ON HER LEG. THE EMPLOYEE RECEIVED MEDICAL TREATMENT FOR A SECOND DEGREE BURN AT THE WORKERS COMPENSATION MEDICAL DEPARTMENT. THE USER FACILITY REPORTED THE EMPLOYEE IS FINE AND HAS NO SUSTAINING INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STEAM GENERATOR STEAM GENERATOR - WASHER COMPONENT FLE STERIS MEXICO, S. DE R.L. DE C.V. CH08-861-500

Patients

Seq Age Sex Outcome Treatment
1 Other