FDA Adverse Event Malfunction Summary report: N

VITALITY

MDR report key: 2042482 · Received April 6, 2011

Report

Report Number
2124215-2011-01369
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
December 9, 2008
Report Date
February 27, 2015
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT PROCEDURE WAS SCHEDULED FOR A LATER DATE. THE AVAILABLE INFORMATION SUGGESTS THIS DEVICE REMAINS IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, TECHNICIANS USED AN ENGINEERING-LEVEL LONGEVITY PREDICTION CALCULATION TO ASSESS THE RATE OF BATTERY DEPLETION, GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE. THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A COMPREHENSIVE SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF THERAPY DELIVERY, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED INDICATING THIS DEVICE WAS LATER EXPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) DUE TO EXTENDED CHARGE TIMES EXCEEDING THE EXTENDED CHARGE TIME LIMIT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T135

Patients

Seq Age Sex Outcome Treatment
1 66 YR T135| 0158