FDA Adverse Event Malfunction Summary report: N

SIGMA MODEL SPECTRUM

MDR report key: 2042414 · Received April 6, 2011

Report

Report Number
1314492-2011-00011
Event Type
Malfunction
Date Received
April 6, 2011
Report Date
April 6, 2011
Manufacturer
SIGMA LLC
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

PRELIMINARY EVAL INDICATES A LOWER BEARING FAILURE OCCURRED. A F/U REPORT WILL BE SUBMITTED, ONCE A FULL EVAL HAS BEEN CONDUCTED BY SIGMA ENGINEERING.

Description of Event or Problem · 1

FAILED FLOW RATE TEST HIGH, RATE 200, VTBI 50, WHEN VOLUME GIVEN INDICATED 10 ML HAD INFUSED THE GRADUATE ALREADY HAD COLLECTED 18 ML. PUMP WAS THEN PUT INTO THE IN-STOP MODE WHERE IT WAS FOUND TO BE FREE FLOWING. THE SYMPTOM WAS DISCOVERED DURING A ROUTINE INSPECTION. CLINICAL ENGINEERING ADDED THAT THEIR FIRST FINDING WAS THAT THE PUMP FAILED THE FLOW RATE TESTS LOW, RATE 400, VTBI 40, DELIVERED 35 ML. CLINICAL ENGINEERING REPORTED THAT THERE WAS NOT ANY INACCURACY OR INCIDENT REPORT FROM NURSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA MODEL SPECTRUM SPECTRUM INFUSION PUMP FRN SIGMA LLC 35700 732239

Patients

Seq Age Sex Outcome Treatment
1