ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2011-03863
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- January 18, 2011
- Report Date
- January 18, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT IN OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL OBSERVATIONS INDICATED A COMPLETE LEAD WAS RETURNED. A KINKED STYLET WAS RETURNED INSIDE THE LEAD. THE TERMINAL TOOL WAS RETURNED ON THE LEAD, AND WAS SEATED CORRECTLY WITH THE END CAP ENGAGED. THERE WERE SET SCREW MARKS NOTED ON THE RATE/SENSE TERMINAL PINS. THE LEAD BODY WAS TWISTED SLIGHTLY ALONG ITS ENTIRE LENGTH, AND THERE WAS BLOOD/BODY FLUID NOTED ON THE TERMINAL PIN. THIS LEAD PASSED THE CONTINUITY TEST WITH MANIPULATION. THE INITIAL ALLEGATION OF HIGH LEAD IMPEDANCE COULD NOT BE CONFIRMED BY ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD, DURING IMPLANT, EXHIBITED OUT OF RANGE PACING IMPEDANCE MEASUREMENTS. THE LEAD WAS FIRST TESTED USING A COMPETITOR'S PACING SYSTEM ANALYZER (PSA), WHICH PRODUCED A PACING IMPEDANCE OF 1,3450 OHMS, AND A THRESHOLD OF 1.3 VOLTS. THE PHYSICIAN ACCEPTED THESE VALUES, AND CONNECTED THIS RV LEAD TO THE ASSOCIATED DEVICE. MEASUREMENTS WITH THE ASSOCIATED DEVICE WERE THEN TAKEN, AND PRODUCED A PACING IMPEDANCE MEASUREMENT OF GREATER THAN 2, 000 OHMS. THE SHOCK AND SENSING VALUES WERE NORMAL AND WITHIN RANGE. THE PHYSICIAN TRIED CLEANING THE LEAD CONNECTOR, BUT THE SAME IMPEDANCE MEASUREMENTS WERE OBSERVED. THIS RV LEAD WAS THEN RECONNECTED TO THE COMPETITOR'S PSA, AND PACING IMPEDANCE MEASUREMENTS OF GREATER THAN 4,000 OHMS WERE OBSERVED. THE DECISION WAS MADE TO IMPLANT A NEW RV LEAD. THIS WAS DONE SUCCESSFULLY, AND ALL PARAMETERS WERE WITHIN NORMAL RANGE. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | NVY | CPI - DEL CARIBE | 0296 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |