FDA Adverse Event Malfunction Summary report: N

ENDOTAK RELIANCE

MDR report key: 2042306 · Received April 6, 2011

Report

Report Number
2124215-2011-03863
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
January 18, 2011
Report Date
January 18, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT IN OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL OBSERVATIONS INDICATED A COMPLETE LEAD WAS RETURNED. A KINKED STYLET WAS RETURNED INSIDE THE LEAD. THE TERMINAL TOOL WAS RETURNED ON THE LEAD, AND WAS SEATED CORRECTLY WITH THE END CAP ENGAGED. THERE WERE SET SCREW MARKS NOTED ON THE RATE/SENSE TERMINAL PINS. THE LEAD BODY WAS TWISTED SLIGHTLY ALONG ITS ENTIRE LENGTH, AND THERE WAS BLOOD/BODY FLUID NOTED ON THE TERMINAL PIN. THIS LEAD PASSED THE CONTINUITY TEST WITH MANIPULATION. THE INITIAL ALLEGATION OF HIGH LEAD IMPEDANCE COULD NOT BE CONFIRMED BY ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD, DURING IMPLANT, EXHIBITED OUT OF RANGE PACING IMPEDANCE MEASUREMENTS. THE LEAD WAS FIRST TESTED USING A COMPETITOR'S PACING SYSTEM ANALYZER (PSA), WHICH PRODUCED A PACING IMPEDANCE OF 1,3450 OHMS, AND A THRESHOLD OF 1.3 VOLTS. THE PHYSICIAN ACCEPTED THESE VALUES, AND CONNECTED THIS RV LEAD TO THE ASSOCIATED DEVICE. MEASUREMENTS WITH THE ASSOCIATED DEVICE WERE THEN TAKEN, AND PRODUCED A PACING IMPEDANCE MEASUREMENT OF GREATER THAN 2, 000 OHMS. THE SHOCK AND SENSING VALUES WERE NORMAL AND WITHIN RANGE. THE PHYSICIAN TRIED CLEANING THE LEAD CONNECTOR, BUT THE SAME IMPEDANCE MEASUREMENTS WERE OBSERVED. THIS RV LEAD WAS THEN RECONNECTED TO THE COMPETITOR'S PSA, AND PACING IMPEDANCE MEASUREMENTS OF GREATER THAN 4,000 OHMS WERE OBSERVED. THE DECISION WAS MADE TO IMPLANT A NEW RV LEAD. THIS WAS DONE SUCCESSFULLY, AND ALL PARAMETERS WERE WITHIN NORMAL RANGE. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE NVY CPI - DEL CARIBE 0296

Patients

Seq Age Sex Outcome Treatment
1