CONTAK RENEWAL
Report
- Report Number
- 2124215-2011-01753
- Event Type
- Injury
- Date Received
- April 6, 2011
- Date of Event
- January 19, 2011
- Report Date
- March 21, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Removal / Correction Number
- Z-0047-08 TO Z-0053-08
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CHARGE TIMES GREATER THAN THE CHARGE TIME LIMIT. END OF LIFE (EOL) WAS DECLARED FOLLOWING A SINGLE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE AND THE RESULTS INDICATED THE MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE AND PROGRAMMED SETTINGS. IT WAS CONCLUDED THAT THIS DEVICE DID NOT EXPERIENCE PREMATURE BATTERY DEPLETION. RATHER, ERI WAS REACHED EARLIER THAN EXPECTED DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THE DEVICE WAS CAPABLE OF DETECTING AND TREATING ARRHYTHMIAS, AS THE BATTERY ITSELF HAD SUFFICIENT CAPACITY REMAINING TO PROVIDE THERAPY IF NEEDED.
AS OF TODAY, NO ADDITIONAL INFORMATION IS AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY-DEFIBRILLATOR DECLARED A BATTERY STATUS OF END OF LIFE (EOL) DUE TO AN EXTENDED CHARGE TIME. THE DEVICE HAS BEEN EXPLANTED. THERE WERE NO ADVERSE PATIENT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Life Threatening| R | T165| 4087| 0185| H177 |