AUTOSOFT 90
Report
- Report Number
- 3003442380-2024-28835
- Event Type
- Malfunction
- Date Received
- October 10, 2024
- Date of Event
- August 29, 2024
- Report Date
- November 26, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244018303
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. PER REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DHR REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DHR REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6000723, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6000723 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 103 AND MANUFACTURED IN THE LINE L8 ON 15-APR-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. THE SUB-ASSEMBLY: GLUING OF TUBING LOT 3C06789 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 55 AND MANUFACTURED IN THE MACHINE SPOT08, ON 30-MAR-2023, WITH A TOTAL OF (B)(4) UNITS. LOT 3C03740 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 55 AND MANUFACTURED IN THE MACHINE SC07, ON 16-MAR-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN ITALY. IT WAS REPORTED THAT THE PATIENT FACED AN OCCLUSION EVENT WITH AN INFUSION SET AND WAS ALERTED BY ALARM 2 FOLLOWED BY ALARM 26. THE BLOCKAGE WAS FOUND TO BE LOCATED IN THE TUBING. PATIENT CHANGED SUPPLIES AS NECESSARY AND RESUMED INSULIN THERAPY. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 354148 | AUTOSOFT 90 | UNO INSET II 60/6 BLUE TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1002823 | 6000723 | 05705244018303 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |