VITALITY
Report
- Report Number
- 2124215-2011-01331
- Event Type
- Injury
- Date Received
- April 6, 2011
- Date of Event
- January 19, 2011
- Report Date
- October 25, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
TO DATE, INFORMATION SUGGESTS THAT THIS MEDICAL DEVICE REMAINS ACTIVELY IN SERVICE. INVESTIGATION IS CONSIDERED OPEN AT THIS TIME. AS NEW INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE FURTHER EVALUATED AND UPDATED.
THE DEVICE WAS EXPLANTED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT WILL BE UPDATED IF MORE INFORMATION BECOMES AVAILABLE.
ADDITIONAL INFORMATION WAS RECIEVED THAT THIS DEVICE WAS EXPLANTED AND REPLACED DUE TO BATTERY DEPLETION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DISPLAYED A MONITORING VOLTAGE MEASUREMENT OF 2.58 VOLTS AND A CHARGE TIME MEASUREMENT OF 20.2 SECONDS. THERE WAS A CONCERN THAT THE BATTERY DEPLETED EARLIER THAN EXPECTED AS A RESULT OF UNEXPECTED CHARGE TIME. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS CLINICAL OBSERVATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Hospitalization| L| R | 0185| 4470| T180 |