FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II 24GAX0.75IN PRN SLM NPVC

MDR report key: 20422549 · Received October 10, 2024

Report

Report Number
3002601200-2024-00517
Event Type
Malfunction
Date Received
October 10, 2024
Date of Event
September 15, 2024
Report Date
October 23, 2024
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830787
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

1. THE CUSTOMER RETURNS A PHOTO, WHICH SHOWS BLOOD OOZING FROM THE END OF THE SEPTUM. 2. DHR/BHR REVIEW LOT#4052017 1-THE PRODUCTS OF THIS BATCH WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN MARCH 2024, AND PACKAGED AT R240 PACKAGE LINE IN MARCH 2024. WORK ORDER QUANTITY WAS (B)(4). EA. 2-REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MET THE PRODUCT SPECIFICATIONS. 3-REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 4-THE PLANT HAS LAUNCHED CAPA TO INVESTIGATE THE LEAKAGE AT THE SEPTUM. 3. AS THE PLANT RECEIVED SIMILAR COMPLAINTS FROM OTHER HOSPITALS ABOUT THIS BATCH OF PRODUCTS, 800MM SIMULATED CLINICAL LEAKAGE TEST WAS CONDUCTED ON THE RETAINED SAMPLES OF THIS BATCH, AND IT WAS FOUND THAT A FEW SAMPLES LEAKED AT THE END OF THE SEPTUM AFTER THE NEEDLE TIP WAS INSERTED INTO THE TEST DEVICE, AND THE LEAKAGE STOPPED AFTER THE NEEDLE CORE WAS PULLED OUT. CONCLUSION(S): NO ABNORMALITY IS FOUND IN THE PRODUCTION PROCESS. THE RETURNED PHOTO SHOWS BLOOD OOZING FROM THE END OF THE SEPTUM. IN RESPONSE TO THIS DEFECT, THE PLANT HAS LAUNCHED CAPA TO TRACE AND INVESTIGATE ITS ROOT CAUSE.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II 24GAX0.75IN PRN SLM NPVC LEAKAGE AT SEPTUM AFTER THE PUNCTURE, OBSERVE FOR AIR AND BLOOD LEAKAGE AT THE STOPPER DURING BLOOD COLLECTION. THE SAMPLE CANNOT BE RETURNED, THERE IS A PHOTO. NO GREEN CLAIM FORM REQUIRED, A RESPONSE TO THE COMPLAINT AND A LETTER ACKNOWLEDGING RECEIPT OF THE COMPLAINT REQUIRED.