FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2042164 · Received March 28, 2011

Report

Report Number
2531779-2011-02038
Event Type
Injury
Date Received
March 28, 2011
Date of Event
February 17, 2011
Report Date
February 26, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Removal / Correction Number
2531779-02/25/11-001-R
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. ALTHOUGH THE CARTRIDGE WAS NOT RETURNED THERE IS NO FURTHER INVESTIGATION NECESSARY WITH RESPECT TO THE LEAK ISSUE. CARTRIDGES WITH LOT# B201582 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.

Description of Event or Problem · 1

THE REPORTER, THE PT'S MOTHER, REPORTED THAT SINCE (B)(6) 2010 THE PT OBTAINED BLOOD GLUCOSE READINGS RANGING FROM 300 MG/DL TO 500 MG/DL. ON (B)(6) 2011 THE PT OBTAINED THE BLOOD GLUCOSE READING OF 550 MG/DL. THE PT EXPERIENCED THE SYMPTOMS OF LETHARGY, DIZZINESS AND HEADACHE. THE PT REC'D A CORRECTION VIA THE PUMP AND TWO HOURS LATER HIS BLOOD GLUCOSE LEVEL WAS 300 MG/DL. ON THAT DATE, THE PT'S MOTHER NOTED THE INSULIN CARTRIDGE WAS LEAKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP CARTRIDGE LZG ANIMAS CORPORATION IR 1200 / 1250 / 2020/ OTP B201582

Patients

Seq Age Sex Outcome Treatment
1 12 YR Life Threatening| R