FDA Adverse Event
Injury
Summary report: N
ANIMAS INSULIN CARTRIDGE
MDR report key: 2042164
·
Received March 28, 2011
Report
- Report Number
- 2531779-2011-02038
- Event Type
- Injury
- Date Received
- March 28, 2011
- Date of Event
- February 17, 2011
- Report Date
- February 26, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Removal / Correction Number
- 2531779-02/25/11-001-R
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. ALTHOUGH THE CARTRIDGE WAS NOT RETURNED THERE IS NO FURTHER INVESTIGATION NECESSARY WITH RESPECT TO THE LEAK ISSUE. CARTRIDGES WITH LOT# B201582 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.
Description of Event or Problem · 1
THE REPORTER, THE PT'S MOTHER, REPORTED THAT SINCE (B)(6) 2010 THE PT OBTAINED BLOOD GLUCOSE READINGS RANGING FROM 300 MG/DL TO 500 MG/DL. ON (B)(6) 2011 THE PT OBTAINED THE BLOOD GLUCOSE READING OF 550 MG/DL. THE PT EXPERIENCED THE SYMPTOMS OF LETHARGY, DIZZINESS AND HEADACHE. THE PT REC'D A CORRECTION VIA THE PUMP AND TWO HOURS LATER HIS BLOOD GLUCOSE LEVEL WAS 300 MG/DL. ON THAT DATE, THE PT'S MOTHER NOTED THE INSULIN CARTRIDGE WAS LEAKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP CARTRIDGE | LZG | ANIMAS CORPORATION | IR 1200 / 1250 / 2020/ OTP | B201582 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR | Life Threatening| R |