LIGAMAX-5MM
Report
- Report Number
- 3005075853-2011-01376
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- January 1, 2011
- Report Date
- February 4, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. LOCKOUT THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED WITH THE JAWS IN THE CLOSED POSITION AND WITHOUT ANY CLIP INSIDE THE JAWS. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT HAD BEEN FIRED THROUGH. PLEASE NOTE THE DEVICE CONTAINS A LOCKOUT FEATURE THAT IS DESIGNED TO INCREASE THE REQUIRED FORCE IT TAKES TO CLOSE THE TRIGGER ONCE THE LAST CLIP HAS BEEN FIRED; THIS REDUCES THE POSSIBILITY OF EMPTY JAWS BEING CLOSED ON A STRUCTURE. IN ADDITION, IF THE TRIGGER IS FORCED CLOSED ONCE THE LOCKOUT HAS BEEN ENGAGED, THE JAWS MAY REMAIN IN THE CLOSED POSITION. FURTHER EVALUATION FOUND THAT ONE JAW WAS BROKEN AT BIFURCATION. THE MOST LIKELY REASON FOR JAW BIFURCATION BREAKAGE IS STRESS CORROSION CRACKING AND THE MOST LIKELY ROOT CAUSE IS EXPOSURE TO A SOLUTION CONTAINING CHLORINE. THIS CONDITION IS VERY UNLIKELY TO OCCUR DURING A SURGICAL PROCEDURE AND IS MOST LIKELY TO OCCUR AFTER POST-SURGERY DEVICE CLEANING. THE JAW BREAKAGE IS NOT OCCURRING AS A RESULT OF THE PRODUCT COMPLAINT DECONTAMINATION PROCESS. THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE JAMMED DURING AN UNKNOWN FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | G4UH3Y |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |