LIGAMAX-5MM
Report
- Report Number
- 3005075853-2011-01375
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- January 1, 2011
- Report Date
- February 8, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. ADVANCER. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE TIP OF THE ADVANCER SLID TOWARD TISSUE STOP DURING ONE FIRING SEQUENCE CAUSING THAT THE JAWS REMAINED IN THE CLOSED. THE TRIGGER WAS MANUALLY ASSISTED IN ORDER TO OPEN THE JAWS WITHOUT ANY DIFFICULTIES NOTED; A PEAR SHAPED CLIP WAS RELEASED. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS NOT VISIBLE. THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE THE DEVICE JAMMED WHEN THE SURGEON ATTEMPTED TO DRY FIRE IT. THEY COMPLETED THE PROCEDURE WITH A NEW LIKE DEVICE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | G4U596 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |