DEXTRUS 4135
Report
- Report Number
- 1028232-2011-00632
- Event Type
- Injury
- Date Received
- March 29, 2011
- Date of Event
- January 28, 2011
- Report Date
- March 14, 2011
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVZ
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS, THEREFORE, BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.
BOSTON SCIENTIFIC REC'D INFORMATION THAT THIS RIGHT ATRIAL LEAD DISLODGED AND FELL INTO THE VENTRICLE. THE LEAD WAS REVISED AND IS NOW WORKING APPROPRIATELY. NO ADVERSE PT EFFECTS WERE REPORTED. AT THIS TIME, THE PRODUCT REMAINS IN SERVICE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED AS NECESSARY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEXTRUS 4135 | PACER LEAD | NVZ | BIOTRONIK SE & CO. KG | 358744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |