FDA Adverse Event Injury Summary report: N

4 FR S/L POLY RADPICC WITH 70 CM PLATINUM TIP GUIDE

MDR report key: 2042015 · Received March 31, 2011

Report

Report Number
3006260740-2011-00094
Event Type
Injury
Date Received
March 31, 2011
Date of Event
March 4, 2011
Report Date
March 7, 2011
Manufacturer
C. R. BARD INC. (BASD)
Product Code
LJS
PMA / PMN Number
K034019
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT OF THE CATHETER BEING OCCLUDED (THROMBOSIS) IS INCONCLUSIVE. RETURNED FOR EVAL ARE SIX SAME LOT NUMBER UN-OPENED 4 FR S/L POLY RADPICC TRAYS. THE COMPLAINT INCIDENT COULD NOT BE CONFIRMED, BASED ON SIX UN-OPENED RADPICC KITS. THIS TYPE OF CLINICAL SITUATION CANNOT BE REPLICATED IN THE LAB. THE PRODUCTS INSTRUCTIONS FOR USE (IFU) INDICATES THAT VENOUS THROMBOSIS IS A POTENTIAL COMPLICATION. A CHR OF LOT #REUL0108 SHOWED NO OTHER SIMILAR PRODUCT COMPLAINT(S) FROM THIS LOT NUMBER.

Description of Event or Problem · 1

VESSEL THROMBOSIS OCCURRED WITHIN 24 HOURS PLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 4 FR S/L POLY RADPICC WITH 70 CM PLATINUM TIP GUIDE LJS C. R. BARD INC. (BASD) REUL0108

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention