Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE AFTER THE FIRST CLIP WAS FIRED, THE DEVICE FIRED A SECOND CLIP AND THEN WAS STUCK ON THE CYSTIC DUCT AND WAS DIFFICULT TO OPEN. THEY MANUALLY OPENED THE DEVICE AND CONTINUED WITH A SECOND DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.