FDA Adverse Event Malfunction Summary report: N

CLIP APPLIER - MEDIUM

MDR report key: 2041971 · Received April 6, 2011

Report

Report Number
1527736-2011-00087
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 10, 2011
Report Date
March 11, 2011
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
GDO
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE JAWS MISALIGNED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE LOADED, RETAINED AND FORMED FIVE CLIPS CONFORMING; HOWEVER, DUE TO THE CONDITION OF THE JAWS ONE CLIP WITH GAP WAS FORMED. A POSSIBLE CAUSE OF THIS CONDITION MAY BE THE INCORRECT/EXCESSIVE APPLICATION OF TORQUE TO THE JAWS DURING INSTRUMENT USE CREATES A MISALIGNMENT OF THE TIPS WHICH MAY RESULT IN CLIP MALFORMATION. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE CLIP APPLIER JAW WAS NOT HOLDING CLIPS PROPERLY, RESULTING IN IMPROPER CLOSURE OF CLIPS. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CLIP APPLIER - MEDIUM CLIP APPLIER - MEDIUM GDO ETHICON ENDO SURGERY, INC (CINCINNATI) NA UNK

Patients

Seq Age Sex Outcome Treatment
1