ACRYSOF
Report
- Report Number
- 1119421-2011-00347
- Event Type
- Injury
- Date Received
- March 30, 2011
- Date of Event
- February 28, 2011
- Report Date
- February 28, 2011
- Manufacturer
- ALCON RESEARCH, LTD./HUNTINGTON
- Product Code
- HQL
- PMA / PMN Number
- P930014
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVAL: THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. STERILIZATION RECORDS FOR THE ASSOCIATED IOL WERE REVIEWED AND THE STERILIZATION CYCLE RAN WITHIN ALL REQUIRED PARAMETERS WITH NO ANOMALIES. RESULTS FROM THE PRODUCT HISTORY RECORD REVIEW INDICATED THE PRODUCT MET RELEASED CRITERIA. ROOT CAUSE: NO SAMPLE WAS RETURNED BY THE CUSTOMER. NO ROOT CAUSE CAN BE DETERMINED AT THIS TIME. ACTION TAKEN: NO FURTHER ACTION IS WARRANTED AT THIS TIME. COMPLAINT TRENDS WILL CONTINUE TO BE MONITORED AND FURTHER ACTION WILL BE INITIATED WHEN DEEMED NECESSARY BY THE APPROPRIATE RESPONSIBLE MANAGEMENT PERSONNEL. THERE HAVE BEEN NO OTHER COMPLAINTS REPORTED IN THE LOT NUMBER. ADDITIONAL INFO WAS REQUESTED ON 03/02/2011 BY FAX AND MAIL. ADDITIONAL INFO WAS RECEIVED IN A FOLLOW UP PHONE CALL ON 03/14/2011. A COMPLETED QUESTIONNAIRE WAS RECEIVED ON 03/17/2011. (B)(4).
A NURSE REPORTED THAT FIVE PTS PRESENTED WITH TOXIC ANTERIOR SEGMENT SYNDROME FIVE DAYS FOLLOWING INTRAOCULAR LENS IMPLANT SURGERY. ALL PTS WERE TREATED WITH MEDICATIONS. NO CULTURES WERE TAKEN. IN A FOLLOW UP, THE SURGEON REPORTED THE IOL DID NOT CAUSE OR CONTRIBUTE TO THE EVENT. THE EVENT HAS NOW RESOLVED. THERE ARE FIVE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS REPORT IS FOR THE FOURTH OF FIVE PTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACRYSOF | INTRAOCULAR LENS | HQL | ALCON RESEARCH, LTD./HUNTINGTON | SN60WF | 10987903 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Required Intervention | BACITRACIN| BSS WITH EPINEPHRINE| MONARCH "D" CARTRIDGE| MIOSTAT| VISCOELASTIC| VIGAMOX| BSS |