POWERPICC SOLO CATHETER WITH SHERLOCK 3CG (TPS) STYLET 4F FULL TRAY
Report
- Report Number
- 3006260740-2024-05849
- Event Type
- Malfunction
- Date Received
- October 10, 2024
- Date of Event
- September 11, 2024
- Report Date
- January 27, 2025
- Manufacturer
- C.R. BARD, INC. (BASD) -3006260740
- Product Code
- LJS
- PMA / PMN Number
- K091324
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. THE COMPLAINT OF A LEAK IS CONFIRMED BUT THE CAUSE IS UNKNOWN. THE PRODUCT RETURNED FOR EVALUATION WAS A 4FR SL POWERPICC SOLO CATHETER. THE RETURNED PRODUCT SAMPLE WAS EVALUATED AND A KINK IMPRESSION WITH A LONGITUDINAL SPLIT WAS OBSERVED BETWEEN THE 15 CM AND 16 CM DEPTH MARKERS ON THE CATHETER BODY. NO SPECIFIC EVIDENCE WAS SEEN DURING THE INVESTIGATION WHICH SUPPORTED A SPECIFIC ROOT CAUSE FOR THIS EVENT. THE DAMAGE LOCATION SUGGESTED THAT CATHETER SECUREMENT, ACCESS, AND/OR MAINTENANCE TECHNIQUES MAY HAVE CONTRIBUTED. THIS COMPLAINT WILL BE RECORDED FOR FUTURE TRENDING AND MONITORING PURPOSES.
THE MANUFACTURER HAS RECEIVED THE SAMPLE AND WILL BE EVALUATED. A SUPPLEMENTAL WILL BE SUBMITTED WITH EVALUATION RESULTS.
IT WAS REPORTED THAT ONCOLOGY PATIENT RECEIVING 3 CONSECUTIVE DAYS OF TREATMENT. DAY 3 WHEN FLUSHED PICC NOTED IMMEDIATE LEAKING AT INSERTION SITE. NO PROBLEM PREVIOUS DAYS. REPEAT FLUSH, SAME OCCURRING. PICC REMOVED. HOLE NOTED IN CATHETER BETWEEN 5-6 CM MARK. CHEMO DELAY. PERIPHERAL ACCESS ESTABLISHED. PICC INITIALLY INSERTED ON (B)(6) 2024 AND NO ISSUES NOTED. R BASILIC 36/40 = OAJ. NO OTHER INFORMATION WAS PROVIDED.
IT WAS REPORTED THAT ONCOLOGY PATIENT RECEIVING 3 CONSECUTIVE DAYS OF TREATMENT. DAY 3 WHEN FLUSHED PICC NOTED IMMEDIATE LEAKING AT INSERTION SITE. NO PROBLEM PREVIOUS DAYS. REPEAT FLUSH, SAME OCCURRING. PICC REMOVED. HOLE NOTED IN CATHETER BETWEEN 5-6 CM MARK. CHEMO DELAY. PERIPHERAL ACCESS ESTABLISHED. PICC INITIALLY INSERTED ON (B)(6) 2024 AND NO ISSUES NOTED. R BASILIC 36/40 = OAJ. NO OTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1933348 | POWERPICC SOLO CATHETER WITH SHERLOCK 3CG (TPS) STYLET 4F FULL TRAY | CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS | LJS | C.R. BARD, INC. (BASD) -3006260740 | N/A | REJQ1580 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | Other |