DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2011-00033
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- March 19, 2011
- Report Date
- March 22, 2011
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD
- Product Code
- JFP
- PMA / PMN Number
- K051089
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED CALCIUM RESULT WAS ALIQUOTTER PROBE MAINTENANCE ISSUES. A SIEMENS HEALTHCARE DIAGNOSTICS INC. FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE SITE AND PERFORMED MAINTENANCE PROCEDURES. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
FURTHER ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED CALCIUM RESULT WAS UNKNOWN. AN INSUFFICIENT SAMPLE IS SUSPECTED. A SIEMENS HEALTHCARE DIAGNOSTICS INC. FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE SITE. NO MALFUNCTIONS WERE DETECTED. THE FSE PERFORMED MAINTENANCE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY DEPRESSED CALCIUM RESULT WAS OBTAINED ON A PATIENT SAMPLE. SEVERAL OTHER ANALYTES ON THE SAME SAMPLE WERE DEPRESSED. THE RESULTS WERE REPORTED TO THE PHYSICIAN. A NEW (REDRAW) SAMPLE WAS RUN. A HIGHER RESULT WITHIN THE NORMAL RANGE WAS OBTAINED FOR THE CALCIUM AND REPORTED. OTHER ANALYTES WERE NOT REPEATED. NO OTHER PATIENT SAMPLES WERE IMPACTED. THE PATIENT RECEIVED A IV INFUSION BASED ON THE ORIGINAL CALCIUM RESULT. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED CALCIUM RESULT.
A FALSELY DEPRESSED CALCIUM RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. A NEW (REDRAW) SAMPLE WAS RUN. A HIGHER RESULT WITHIN THE NORMAL RANGE WAS OBTAINED AND REPORTED. THE PATIENT RECEIVED A IV INFUSION BASED ON THE ORIGINAL CALCIUM RESULT. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED CALCIUM RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | JFP | SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD | VISTA 1500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |