FDA Adverse Event Malfunction Summary report: N

DIMENSION® CLINICAL CHEMISTRY SYSTEM

MDR report key: 2041892 · Received April 6, 2011

Report

Report Number
1226181-2011-00032
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 23, 2011
Report Date
March 24, 2011
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD
Product Code
MMI
PMA / PMN Number
K944093
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS HM CHROME CONTAMINATION. A SIEMENS HEALTHCARE DIAGNOSTICS INC. TECHNICAL SERVICE CENTER (TSC) REPRESENTATIVE DIRECTED THE CUSTOMER TO MAINTENANCE PROCEDURES. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED AND A LOWER RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION® CLINICAL CHEMISTRY SYSTEM CLINICAL CHEMISTRY SYSTEM MMI SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD RXL MAX HM