FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 2041831 · Received April 6, 2011

Report

Report Number
2124215-2011-01579
Event Type
Injury
Date Received
April 6, 2011
Date of Event
January 12, 2011
Report Date
January 12, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME THE LEAD HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC, CRM FOR ANALYSIS. THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF FURTHER INFORMATION BECOMES AVAILABLE THIS EVENT WILL BE REOPENED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE LEFT VENTRICULAR (LV) LEAD WAS EXHIBITING HIGH OUT OF RANGE IMPEDANCE MEASUREMENTS GREATER THAN 2,500 OHMS. THERE WAS ALSO NO CAPTURE. THE LEAD WAS EXPLANTED AND SUCCESSFULLY REPLACED. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4543

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention (B)(4)| (B)(4)| (B)(4)| (B)(4)