FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 2041817 · Received April 6, 2011

Report

Report Number
2124215-2011-02909
Event Type
Injury
Date Received
April 6, 2011
Date of Event
January 13, 2011
Report Date
February 25, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE COMPLETE LEAD RETURNED AND UNDERWENT DETAILED ANALYSIS. VISUAL INSPECTION NOTED THAT A VERY KINKED UNKNOWN GREY HANDLED STYLET WAS RETURNED INSERTED IN THE LEAD. THE PROXIMAL END OF THE DISTAL SPRING ELECTRODE WAS SEPARATED FROM THE LEAD BODY INSULATION WITH MEDICAL ADHESIVE NOTED. SETSCREW MARKS WERE NOTED ON THE IS-1 TERMINAL PIN AND DISTAL HIGH VOLTAGE TERMINAL PIN. NO DISCERNABLE SETSCREW MARKS WERE NOTED ON THE IS-1 RING. THERE APPEARED TO BE SOME TYPE OF HEAT DAMAGE, MOST LIKELY FROM CAUTERY OR LASER FROM THE EXTRACTION. THE GORE WAS DAMAGED, BLACKENED AND MELTED IN MANY PLACES. HELIX WAS SLIGHTLY BENT. THERE WAS BLOOD/BODY FLUID NOTED IN THE LUMENS AND THE HELIX HOUSING. MICROSCOPIC ANALYSIS REVEALED THE TRILUMEN INSULATION DAMAGE THROUGH TO THE DISTAL HIGH VOLTAGE (HV) LUMEN 110-125 MM FROM THE TERMINAL PIN. THE DAMAGE APPEARED AS A SMOOTH SURFACE ABRASION WITH JAGGED EDGES. THERE WAS WEBBING DAMAGE NOTED THROUGH TO THE RS- LUMEN IN THIS AREA AS WELL. THE DISTAL HV CABLE GORE WAS TORN/ABRADED IN THIS AREA. THE DISTAL HV CABLE CONDUCTOR HAD BEEN RUBBED AND APPEARED SLIGHTLY FLATTENED. DETAILED ANALYSIS UNABLE TO CONFIRM THE LEAD FRACTURE. DUE TO THE LOCATION AND TYPE OF DAMAGE IT WAS LIKELY CAUSED BY LEAD ON CAN CONTACT, COUPLED WITH PRESSURE.

Additional Manufacturer Narrative · 1

THE LEAD WAS RETURNED AND DETAILED ANALYSIS IS PENDING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEAD WAS EXPLANTED AS IT WAS FRACTURED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0180

Patients

Seq Age Sex Outcome Treatment
1 153 MO Required Intervention MISMATCH| E102| 0180