MDR report key: 2041801 · Received April 6, 2011

Report

Report Number
2124215-2011-01017
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
January 13, 2011
Report Date
February 4, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
PMA / PMN Number
P960004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. DRIED BODY FLUID WAS NOTED IN THE HELIX LUMEN. THIS LIKELY WAS INTRODUCED THROUGH THE TERMINAL PIN. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Additional Manufacturer Narrative · 1

THIS LEAD REMAINS IMPLANTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS NOT BEEN RETURNED. IF RETURNED, OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL UPDATED.

Description of Event or Problem · 1

INFORMATION WAS LATER RECEIVED THAT THIS LEAD WAS EXPLANTED DUE TO LOSS OF CAPTURE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS RIGHT ATRIAL LEAD EXHIBITED NO SENSING AND NO THRESHOLD MEASUREMENTS. IT WAS SUSPECTED THAT THE LEAD HAD DISLODGED. AS A RESULT, A REVISION PROCEDURE WAS SCHEDULED. NO ADVERSE PATIENT EFFECTS HAVE BEEN NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4479

Patients

Seq Age Sex Outcome Treatment
1 52 YR Required Intervention S603| 4479| 4456