VITALITY 2

MDR report key: 2041715 · Received April 6, 2011

Report

Report Number
2124215-2011-00988
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
January 13, 2011
Report Date
January 18, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

REVIEW OF DEVICE MEMORY REVEALED THE DEVICE DECLARED ITS ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. ADDITIONAL LAB ANALYSIS AND TESTING SHOWED ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF DEVICE INTERNAL BATTERY IMPEDANCE. THE EOL (END OF LIFE) INDICATOR WAS NOT DECLARED WHILE THE DEVICE WAS IN SERVICE. TESTING CONFIRMED ALL DEVICE THERAPIES WERE AVAILABLE.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS DEVICE'S PROGRAMMED PARAMETER SETTINGS AND THERAPY HISTORY WERE USED TO ESTIMATE EXPECTED BATTERY USE TO DATE, THEN COMPARED TO ACTUAL USE. OUR INITIAL ANALYSIS SHOWED THIS DEVICE DID NOT MEET LONGEVITY EXPECTATIONS PER PROGRAMMED VALUES. ADDITIONAL ANALYSIS IS PENDING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. THERE WERE NO KNOWN ALLEGATIONS AGAINST THE DEVICE, AND NO REPORTS OF ADVERSE PATIENT EFFECTS. THIS DEVICE IS NOT INCLUDED IN ANY PRODUCT ADVISORY POPULATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T175

Patients

Seq Age Sex Outcome Treatment
1 56 YR T175| 0185