VITALITY 2
Report
- Report Number
- 2124215-2011-00988
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- January 13, 2011
- Report Date
- January 18, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
REVIEW OF DEVICE MEMORY REVEALED THE DEVICE DECLARED ITS ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. ADDITIONAL LAB ANALYSIS AND TESTING SHOWED ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF DEVICE INTERNAL BATTERY IMPEDANCE. THE EOL (END OF LIFE) INDICATOR WAS NOT DECLARED WHILE THE DEVICE WAS IN SERVICE. TESTING CONFIRMED ALL DEVICE THERAPIES WERE AVAILABLE.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS DEVICE'S PROGRAMMED PARAMETER SETTINGS AND THERAPY HISTORY WERE USED TO ESTIMATE EXPECTED BATTERY USE TO DATE, THEN COMPARED TO ACTUAL USE. OUR INITIAL ANALYSIS SHOWED THIS DEVICE DID NOT MEET LONGEVITY EXPECTATIONS PER PROGRAMMED VALUES. ADDITIONAL ANALYSIS IS PENDING.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. THERE WERE NO KNOWN ALLEGATIONS AGAINST THE DEVICE, AND NO REPORTS OF ADVERSE PATIENT EFFECTS. THIS DEVICE IS NOT INCLUDED IN ANY PRODUCT ADVISORY POPULATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | T175| 0185 |