FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 2041695 · Received April 6, 2011

Report

Report Number
2124215-2011-01030
Event Type
Injury
Date Received
April 6, 2011
Date of Event
January 13, 2011
Report Date
March 22, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Additional Manufacturer Narrative · 1

AT THIS TIME THIS PRODUCT HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC FOR ANALYSIS. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS TRANSVENOUS DEFIBRILLATION LEAD EXHIBITED HIGH PACING THRESHOLDS AND OUT OF RANGE RIGHT VENTRICULAR (RV) IMPEDANCES OF GREATER THAN 2,000 OHMS ONE DAY POST-OP. IT WAS NOTED THAT A MICRO-PERFORATION WAS SUSPECTED; THEREFORE, A REVISION PROCEDURE WAS PERFORMED. THIS LEAD WAS EXPLANTED AND A NEW LEAD WAS SUCCESSFULLY IMPLANTED. AFTER THE IMPLANT OF THE NEW LEAD, AN ECHOCARDIOGRAM WAS PERFORMED. THERE WAS NO EVIDENCE OF PERFORATION OR CARDIAC EFFUSION PRESENT; THEREFORE, A MICRO-DISLODGEMENT OF THE ORIGINAL LEAD WAS LATER SUSPECTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

THIS LEAD WAS LATER RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0175

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention E110| 0175| 0185| 4470