FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 2041688 · Received April 6, 2011

Report

Report Number
2124215-2011-01443
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
January 15, 2011
Report Date
January 8, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT WAS BROUGHT INTO CLINIC AND HIGH PACING IMPEDANCE MEASUREMENTS WERE STILL OBSERVED. ALL OTHER LEAD DIAGNOSTICS WERE ACCEPTABLE AND THERE WAS NO EVIDENCE OF NOISE. TECHNICAL SERVICES DISCUSSED CONTINUED MONITORING OR REPLACING THE RATE/SENSE PORTION OF THE LEAD. THE AVAILABLE INFORMATION SUGGESTS THIS DEVICE AND LEAD REMAIN IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED INDICATING THIS DEVICE WAS LATER EXPLANTED DUE TO NORMAL BATTERY DEPLETION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT A LATITUDE ALERT WAS ISSUED FOR THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND IMPLANTABLE DEFIBRILLATION LEAD DUE TO HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR LWP GUIDANT CRM CLONMEL IRELAND H219

Patients

Seq Age Sex Outcome Treatment
1 52 YR H219| 4469| 4549| 4518| 0147