FDA Adverse Event Injury Summary report: N

VITALITY

MDR report key: 2041668 · Received April 6, 2011

Report

Report Number
2124215-2011-01146
Event Type
Injury
Date Received
April 6, 2011
Date of Event
January 14, 2011
Report Date
May 27, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

OUR RECORDS INDICATE THAT THIS DEVICE REMAINS IMPLANTED. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

AT THIS TIME, THIS DEVICE HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WERE ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), WHICH IS INCLUDED IN THE MID-LIFE DISPLAY OF REPLACEMENT INDICATOR ADVISORY, TRIGGERED ELECTIVE REPLACEMENT INDICATOR (ERI) WITH A MONITORING VOLTAGE OF 2.54 VOLTS AND CHARGE TIME MEASUREMENTS OF 20 SECONDS. NO ADVERSE PATIENT EFFECTS WERE REPORTED

Description of Event or Problem · 1

ADDITIONAL INFORMATION INDICATED THAT THIS DEVICE WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T135

Patients

Seq Age Sex Outcome Treatment
1 54 YR 0158| T135